ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS)

0 out of 5

Description

TYPE

Type of Atomic Absorption Spectrometer (AAS)Double beam
Monochromator typeCzerny-turner
Detector TypePhoto Multiplier Tube (PMT)

SPECTRAL

Source Wavelength – Minimum180 nanometer
Source Wavelength – Maximum900 nanometer
Accuracy (+/-)0.1 nanometer
Repeatability0.3 nanometer
Diffraction grating (lines/mm)1800
Wavelength setting and scanningAutomatic
Slit width AdjustmentAutomatic
Slit width range(nm)0.1 – 2.0
Resolution (nm/mm)0.1 parts per million
Bandwidth (nm)0.1 – 2.0
Focal length (mm)300 and above

LIGHT SOURCE

Light SourceHollow Cathode Lamp (HCL)
Type of HCLBoth Single and Mulitiple Element
Number of Single element HCLs supplied (inclusive in the scope of supply)4
Number of multiple element (alloy) HCLs supplied (inclusive in the scope of supply)N/A
Measuring modes (K factor and standards)Transmittance,Absorbance,Concentration (Wavelength, time),Emission
Detectable Elements using single element HCLCu,Fe,Zn,Mn,Mg,B,K,Mo,Ca,As
Detectable Elements using multi element HCLNa-K,Ca-Mg,Si-Al,Fe-Ni,Sr-Ba,Al-Ca-Mg,Ca-Mg-Zn,Cu-Mo-Co-Zn,Cd-Cu-Pb-Zn,Cu-Fe-Zn-Mn,Co-Cr-Cu-Fe-Mn-Ni,Cd – Hg – Pb
Measurement level%,ppm level,ppb level
Lifespan of Single element lamp (mA-Hr)5000
Lifespan of multi-element/Alloys lamp (mA-Hr)5000
Number of sockets on Turret8

ATOMIZER

Type of atomizerFlame Burner
NebulizerPt/Ir nebulizer
Background CorrectionD2 Lamp method
Flame and furnace change over switchN/A
Burner diameter (mm)100
CoolantAir type
Sample volume10 milliliter
Corrosion resistant ChamberYes
Adjustable aerosol based nebulizer flowYes
Furnace heating range (Degree celcius)N/A (For burner type)
Time for analysis (sample)10 second

GENERIC

Displaythrough PC
Display typeNA
Test result printingThrough external printer
If inbuilt printer, Type of printerN/A
ConnectivityN/A
Safety sensorsPressure sensor,Flame Sensor,Flame Ignition Sensor,Burner interlock,Power failure protection,Auto ignition
Burner movement (lateral,horizontal & height)Automatic

POWER SOURCE

Power Supply230 V A.C ± 10%,50 Hz

OPERATING CONDITIONS

Minimum Operating Temperature5 degree Celsius
Maximum Operating Temperature50 degree Celsius
Operating Humidity (RH) (%) at 40 degree c95 percent

ACCESSORIES

Type of gas cylindersAcetylene
Type of Regulators suppliedWith Heater
Number of Regulators Supplied (inclusive in the scope of supply)1
Number of Acetylene Cylinders supplied (inclusive in the scope of supply)1
Capacity of each Acetylene cylinder (Litres)47
Number of Nitrous oxide Cylinders supplied (inclusive in the scope of supply)1
Capacity of each Nitrous oxide cylinder (Litres)47
Gas cylinders confirming toIS : 7285 latest
Gas cylinders having PESO certificateYes
Other accessories (inclusive in the scope of supply)Gas flow control box,Oil Moisture Filters,SS Exhaust fume hood with blower,Schematic diagram and instruction manual
Suitable software for connecting PC & Printer (inclusive in the scope of supply)Yes

WARRANTY / AVAILABILITY OF SPARES

warranty (Year)1 year
Spares and Consumables inclusive in the scope of supplyYes

CERTIFICATION

Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use General requirementsYes
Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory useYes
Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specificationYes
If yes test report to furnished,to the buyer on demandYes
Test Report number and dateECIL/ISD/PHOTONICS/AAS/01 & 02.07.2018
Name and address of the LabECIL

Reviews

There are no reviews yet.

Be the first to review “ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS)”