acute care monitoring units and related products-patient monitors

0 out of 5
Category: I.C.U. Equipment

Description

NEW SET

Type of MonitorPreconfigured/Non Modular
Functional Requirements: Operates from mains voltage or from internal rechargeable battery Operator can set audio visual alarm levels for low or high levels of each parameter independently Allows display of single, 3 lead ECG or simultaneous display of at least 5 waves ECG selected from up to 12 points Display to be digital of all active parameters and trace display for at least three selectable parametersYes
Clinical Purpose: Designed to continuously measure and display multiple vital physiological parameters of adults, newborn and premature infants, especially those under critical careYES
Continuous display on screen of neonatal or infant ECG, respiration and heart rates, invasive / non-invasive blood pressure, body temperature and SpO2Yes
Patient monitor offered with option to monitor parameter(s):EtCO2,IBP,Cardiac Output
Number of leads for ECG5
Screen Size12 inch
Weight5 kilogram
ECG patient connectors that are sterilisable and reusable are acceptable though reusable cables that attach to disposable connection patches are preferred /
Multichannel (up to 12 leads) ECG measurement and selectable display of upto 5 leads at a time /
Temperature probe to be reusable, external skin contact type; Temperature range at least 30 to 40 deg C, minimum gradation (1/10) deg C / Heart rate measurement range to be at least 30 to 250 bpm, with accuracy better than ± 5 bpm and minimum gradation 1 bpm /
SpO2 measurement range at least 40-70 % and 70 to 99 %, with accuracy better than ± 1% for 40-70 range and better than ± 3% for 70-99 range and minimum gradation 1% /
Blood pressure monitoring range at least 30 to 300 mmHg, minimum gradation 1 mmHg /
Respiration rate measurement range at least 0 to 100 bpm, minimum gradation 1 bpm /
Trend display of each parameter over at least previous 24 hours to be selectableYes
ECG patient connectors that are sterilisable and reusable are acceptable though reusable cables that attach to disposable connection patches are preferred /
Multichannel (up to 12 leads) ECG measurement and selectable display of upto 5 leads at a time /
Temperature probe to be reusable, external skin contact type; Temperature range at least 30 to 40 deg C, minimum gradation (1/10) deg C / Heart rate measurement range to be at least 30 to 250 bpm, with accuracy better than ± 5 bpm and minimum gradation 1 bpm /
SpO2 measurement range at least 40-70 % and 70 to 99 %, with accuracy better than ± 1% for 40-70 range and better than ± 3% for 70-99 range and minimum gradation 1% /
Blood pressure monitoring range at least 30 to 300 mmHg, minimum gradation 1 mmHg /
Respiration rate measurement range at least 0 to 100 bpm, minimum gradation 1 bpm /
Trend display of each parameter over at least previous 24 hours to be selectable/ TFT ScreenYes
User Interfaceremotely operated
Audio Visual alarmshigh and low levels for each parameter (operator variable settings), sensor / wire / probe disconnected, low battery
Case is to be hard and splash proof / Display must allow easy viewing in all ambient light levels /Cable connectors to be designed so as fit correct socket onlyYes
Noise (in dB)< 50
Heat Dissipation:Should maitain nominal Temp and the heat should be disbursed through adequate arrangement
Mobility & Portabilitymobile and portable in protective case supplied for clean storage and safe transport.
Power supply220 volt, 50 Hz AC
Mains cable to be at least 3m lengthYES
Accessories & Spare Parts:pairs of ECG cable 5 sets of ECG connection electrodes 1 unit each of reusable SPO2 probe(Adult/Paediatric) 1 unit of SPO2 probe(Neo natal) 2 units of NIBP cuffs for adult 2 units of NIBP cuffs for Paediatric 1 unit each of skin and rectal Temperature probe
Operating condition:apable of operating continuously in ambient temperature of 0 to 50 deg C and relative humidity of 15 to 90% in ideal circumstances
Should be FDA / CE/BIS approved productUS FDA,European CE
//- Electrical safety conforms to standards for electrical safety IEC-60601-1 //- Shall meet IEC-60601-1-2 (General requirements for safety – electromagnetic compatibility) //-Also conforms to IEC 60601-2- 27 (Particular requirements for the safety of electrocardiographic monitoring equipment) //- Manufacturer / supplier should have ISO certificate for quality standard //- Shall meet IEC 60601-2-49 multi function patient monitoring equipment standard requirementYes
Seller to perform supply, installation, safety & operation checks and training before handover of the product [Training of users in operation and basic maintenance shall be provided]Yes
WarrantyComprehensive warranty of 3 years with free servicing ( min 3 /year) during warranty
Operating manuals, service manuals, other manuals/documentsAdvanced maintenance tasks required shall be documented
User, technical and maintenance manuals to be supplied in English language.
List to be provided of equipment and procedures required for local calibration and routine maintenance List to be provided of important spares and accessories, with their part numbers. Certificate of calibration and inspection to be provided.
Recommendations or warningsAny recommendations for best use and supplimentary warning for safety to be declared

Details Of Item And Test Reports

Availability of Test Report from Central Govt /NABL/ILAC/Any OPCW accredited lab to prove conformity to specificationNo
Test report (if indicated as available) to be furnished to buyer on demandYES
Test report No & dateNA
Name and address of the Lab:NA